Table 1

Image quality and artefact scores for the 2D reference and 3D 5× k-t SENSE images before and after contrast.


Median Image Quality Score
Median Artefact Score

2D reference images
5
0

3D 5× k-t pre-contrast (p value vs 2D reference score)
3 (p < 0.01)
2 (p < 0.01)

3D 5× k-t post-contrast
(p value vs 2D reference score)
[p value vs pre-contrast 3D score]
4
(p < 0.01)
[p = 0.11]
1
(p < 0.01)
[p = 0.23]

Legend for Table 1

Image quality scoring: (1) endocardial border not visible (non-diagnostic image), (2) severe blurring of endocardial border, (3) moderate blurring, (4) mild blurring, (5) well defined endocardial border (modified from McConnell et al [13])

Artefact scoring: (0) no artefact, (1) minor artefact, (2) significant artefact but not affecting endocardial border definition, (3) significant artefact obscuring endocardial border in places

Maredia et al. Journal of Cardiovascular Magnetic Resonance 2008 10:27   doi:10.1186/1532-429X-10-27